Abstract
Background
To compare image quality, lesion detection and patient comfort of 3T prostate MRI using a combined rigid two-channel phased-array endorectal coil and an external phased-array coil (ERC-PAC) compared to external PAC acquisition in the same patients.
Methods
Thirty three men (mean age 65.3y) with suspected (n = 15) or biopsy-proven prostate cancer (PCa, n = 18) were prospectively enrolled in this exploratory study. 3T prostate MRI including T2-weighted imaging (T2WI) and diffusion-weighted imaging (DWI) was performed using an ERC-PAC versus PAC alone, in random order. Image quality, lesion detection and characterization (biparametric PI-RADSv2.1) were evaluated by 2 independent observers. Estimated signal-to-noise ratio (eSNR) was measured in identified lesions and the peripheral zone (PZ). Patient comfort was assessed using a questionnaire. Data were compared between sequences and acquisitions. Inter/intra-observer agreement for PI-RADS scores was evaluated.
Results
Twenty four prostate lesions (22 PCa) were identified in 20/33 men. Superior image quality was found for ERC-PAC compared to PAC for T2WI for one observer (Obs.1, p < 0.03) and high b-value DWI for both observers (p < 0.05). The sensitivity of PI-RADS for lesion detection for ERC-PAC and PAC acquisitions was 79.2 and 75% for Obs.1, and 79.1 and 66.7%, for Obs.2, without significant difference for each observer (McNemar p-values ≥0.08). Inter−/intra-observer agreement for PI-RADS scores was moderate-to-substantial (kappa = 0.52–0.84). Higher eSNR was observed for lesions and PZ for T2WI and PZ for DWI using ERC-PAC (p < 0.013). Most patients (21/33) reported discomfort at ERC insertion.
Conclusion
Despite improved image quality and eSNR using the rigid ERC-PAC combination, no significant improvement in lesion detection was observed, therefore not supporting the routine use of ERC for prostate MRI.
Δεν υπάρχουν σχόλια:
Δημοσίευση σχολίου