Abstract
Purpose
The objective of this study was to determine whether ultrasound-measured jugular venous pressure (U-JVP) could accurately estimate central venous pressure (CVP).
Methods
This prospective, diagnostic, single-centre study was performed at the Cardiac Intensive Care Unit of the Northern General Hospital, Sheffield, UK. Post-cardiac surgery patients were recruited from January to May 2019. The investigators were blinded to the central venous pressure when measuring the jugular venous pressure. U-JVP and direct CVP were measured simultaneously. Measurements were taken whilst the patient was ventilated and then repeated when the patient was extubated, providing non-ventilated readings.
Results
One-hundred and fourteen consecutive participants with a male predominance of 71% and mean age of 65 ± 12 years were included in the analysis. Bland–Altman plots revealed that U-JVP marginally overestimated CVP by 0.91 mmHg (95% limits of agreement were −2.936 to 4.754) in ventilated patients and by 0.11 mmHg (95% limits of agreement between −2.481 and 2.695) in non-ventilated patients. Reasonable sensitivity and specificity of ultrasound-measured jugular venous pressure was achieved for low and high central venous pressure in both ventilated and non-ventilated patients.
Conclusion
U-JVP accurately estimates cardiac filling pressure and fluid status in patients after cardiac surgery, irrespective of their ventilatory status. Jugular venous pressure measurement by insonation is a reliable technique that can be taught to medical students and other health professionals to non-invasively estimate central venous pressure and may be useful for assessment of volaemic status in patients with heart failure.
Trial registration
ClinicalTrials.gov public (identifier NCT03815188).